Microsoft powerpoint - quant poster white background
Quantitative LC-MS: “How Certain Can we be!”
Peter Stokes, Gavin O’Connor, Chris Mussell, & Ken Webb
LGC Limited, Queens Rd. Teddington, Middlesex, TW11 0LY. INTRODUCTION THE ANALYSIS OF CREATININE IN HUMAN SERUM CREATININE MEASUREMENTS USING DIFFERENT Less than 20% of all LC-MS publications involve MASS SPECTROMETRIC TECHNIQUES quantitation. The reasons for the slow uptake of LC-MS for quantitative analysis have been attributed to poor stability LC-MS Extraction GC-MS extraction of atmospheric pressure ionisation and problems associated with competing matrix reactions. 1. 1g Serum 1. 1g Serum 2. Add Hydrochloric acid (0.1M) However, recent advancements in LC-MS technology such 2. Add labelled creatinine as orthogonal spraying, the use of curtain or cone gases 3. Dilute in water 3. Equilibrate and Z-SprayTM inlets have improved measurement precision 4. Add labelled creatinine and the use of tandem mass spectrometry have helped to 4. Precipitate Protein (3mL ethanol) 5. Equilibrate overcome many problems associated with matrix 5. Centrifuge 15mins, 400rpm 6. Remove creatine using ion interference’s. 6. Evaporate to dryness @ 60°C exchange GC-MS has traditionally been the tool of choice for high 7. Reconstitute in mobile phase 7. Remove creatinine accuracy measurements because measurements typically 8. Run on LC-MS 8. Evaporate to dryness have very low CV’s. 9. Derivatise with MSTFA and Concent 18.5 When comparing GC-MS with LC-MS it is important to pyridine consider the extraction AND measurement process. LC-MS 10. Run on GC-MS methods are often much simpler than GC-MS methods and although the final measurement may be less precise, simplification of the extraction method means that results generated by the two techniques are more comparable than one might think. Extraction time ~ 3 hours Extraction time ~ 7 hours EXACT MATCHING ISOTOPE DILUTION MASS QUANTITATION SUMMARY USING LC-MS WITH MEASUREMENT UNCERTAINTY SPECTROMETRY (EM-IDMS) DIFFERENT CALIBRATION REGEIMES In this form of IDMS samples and standards are prepared The amount of creatinine in the sample is calculated gravimetrically to contain equimolar concentrations of analyte and labelled analogue
W = the concentration of creatinine in sample
W = the concentration of natural creatinine solution used to prepare the calibration blend
m = mass of the natural creatinine standard added to the calibration blend;
This results in both sample and standard being
m = mass of the labelled creatinine standard added to the calibration blend;
indistinguishable on a mass spectrometer
m = mass of the labelled creatinine standard added to the sample blend;
R’ = measured ratio of the sample blend;
= Average measured ratio of the calibration blend injected before and after the sample
Advantages: The measurement uncertainty is defined by: Excellent accuracy and precision Uncertainty equation includes all components of concentration calculation
c = concentration of creatinine in the standard solution
⎝ y ⎠ ⎝ z ⎠ ⎝ yc ⎠
Low uncertainty Concent 18
m = mass of the creatinine-d3 added to the sample blend
Disadvantages:
m = mass of creatinine used to prepare the calibration blend
m = mass of creatinine-d3 added to the calibration blend
Iterative process – need to repeat several times Gravimetric preparation is time consuming Uncertainty budget for the analysis of creatinine using EM-IDMS Requires large amount of instrument time Reduction of Systematic Errors in Quantitative Analysis by Isotope Dilution Mass Spectrometry (IDMS): An Iterative Method; Henrion, A., Fresenius J. Anal. Chem., 1994, 350, 657-658 Guideline for Achieving High Accuracy in Isotope Dilution Mass spectrometry (IDMS); edited by M. Sargent, C. Harrington & R. Harte. Published by RSC (UK) 2002. OTHER HIGH ACCRACY LC-MS(MS) APPLICATIONS Conclusion The results shown in this presentation show that LC-MS Ethynylestradiol in Drinking Water can be an excellent tool for high accuracy applications and Cholesterol in serum* Level 1 (~2ng/g) with the correct choice of calibration regime the results obtained can be comparable with more traditional forms of Lin. cal. solvent standards 1.1ng/g ± 0.66 mass spectrometry such as GC-MS. 1.7840mg/g ± 0.0076 (GC-MS) Lin cal. matrix standards 2.21ng/g ± 1.03 1.7870mg/g ± 0.0059 (LC-MS) Lin. cal. Matrix std’s Ethynylestradiol-d4 IS 2.06ng/g ±0.24 ACKNOWLEDGEMENTS Lin. cal. Matrix std’s Estradiol-d4 IS 2.07ng/g ±0.52 2.7180mg/g ± 0.0056 (GC-MS) Mike Welch - NIST EM-IDMS 1.98ng/g 2.7137mg/g ± 0.0082 (LC-MS) Andre Henrion - PTB Celine S. J. Wolff Briche - LGC Level 2 (~20ng/g) *Comparison of gas chromatography and liquid chromatography mass spectrometric David Carter - LGC measurements for high accuracy analysis of cholesterol in human serum by isotope Lin. cal. solvent standards 23.84ng/g ± 3.10 dilution mass spectrometry. Celine S.J. Wolff Briche. David Carter and Kenneth S. Webb. Rapid Commun. in Mass Spectrom. 2002; 16: 848 - 853. The work described in this poster was Lin cal. matrix standards 13.72ng/g ± 5.78 supported under contract with the Lin. cal. Matrix std’s Ethynylestradiol-d4 IS 19.65ng/g ± 3.53 LSD in Urine Department of Trade and Industry as part of the Valid Analytical Lin. cal. Matrix std’s Estradiol-d4 IS 20.07ng/g ± 3.21 1.568ng/g ± 0.083ng/g Measurement programme. EM-IDMS 19.94ng/g
PROPORTIONERS (3, 4, 6, AND 8 IN.) APPLICATION flow direction arrow, and the type and per-nozzle shall terminate in the foam concen-Ansul proportioners are specifically designedto accurately proportion and control the mix-ing of pressurized Ansul foam concentratesinto a water stream with minimum pressurerounded inlet and a smooth machined finishThe foam concentrate metering orific
ANIMAL BEHAVIOUR, 1998, 56, 1465–1470 Article No. ar980919 Female sexual preferences differ in Mus spicilegus and Mus musculus domesticus : the role of familiarization B. PATRIS & C. BAUDOIN Laboratoire d’Ethologie Expe´rimentale et Compare´e, CNRS UPRES-A 7025, Universite´ Paris-Nord (Received 19 January 1998; initial acceptance 19 February 1998; final acceptance 19 June